B, J, K, R, T Type and C, F Scale TC Select Routine in a MET/CAL Procedure
If you want a way to select the TC type and scale in a procedure, the following procedure example gives one way to do this.
Fluke - San Diego MET/CAL Procedure
=============================================================================
INSTRUMENT: B,J,K,R,T type and C,F scale TC select routine
DATE: 2002-05-01
AUTHOR: Bill Spath
REVISION: 0
ADJUSTMENT THRESHOLD: 70%
NUMBER OF TESTS: 1
NUMBER OF LINES: 191
=============================================================================
STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON
1.001 ASK- R P U F W
1.002 ASK+ X
1.003 LABEL DATA_ENTRY
1.004 MEM2 Enter the TC Scale (C, F, or K)
1.005 MATH S[10] = MEM2
1.006 MEM2 Enter the TC Type (J, K, T, E, N, B, R or S)
1.007 MATH S[11] = MEM2
1.008 JMPL J_TYPE_C ZCMPI(S[11], "J") + ZCMPI(S[10], "C") > 1
1.009 JMPL K_TYPE_C ZCMPI(S[11], "K") + ZCMPI(S[10], "C") > 1
1.010 JMPL T_TYPE_C ZCMPI(S[11], "T") + ZCMPI(S[10], "C") > 1
1.011 JMPL E_TYPE_C ZCMPI(S[11], "E") + ZCMPI(S[10], "C") > 1
1.012 JMPL N_TYPE_C ZCMPI(S[11], "N") + ZCMPI(S[10], "C") > 1
1.013 JMPL B_TYPE_C ZCMPI(S[11], "B") + ZCMPI(S[10], "C") > 1
1.014 JMPL R_TYPE_C ZCMPI(S[11], "R") + ZCMPI(S[10], "C") > 1
1.015 JMPL S_TYPE_C ZCMPI(S[11], "S") + ZCMPI(S[10], "C") > 1
1.016 JMPL J_TYPE_F ZCMPI(S[11], "J") + ZCMPI(S[10], "F") > 1
1.017 JMPL K_TYPE_F ZCMPI(S[11], "K") + ZCMPI(S[10], "F") > 1
1.018 JMPL T_TYPE_F ZCMPI(S[11], "T") + ZCMPI(S[10], "F") > 1
1.019 JMPL E_TYPE_F ZCMPI(S[11], "E") + ZCMPI(S[10], "F") > 1
1.020 JMPL N_TYPE_F ZCMPI(S[11], "N") + ZCMPI(S[10], "F") > 1
1.021 JMPL B_TYPE_F ZCMPI(S[11], "B") + ZCMPI(S[10], "F") > 1
1.022 JMPL R_TYPE_F ZCMPI(S[11], "R") + ZCMPI(S[10], "F") > 1
1.023 JMPL S_TYPE_F ZCMPI(S[11], "S") + ZCMPI(S[10], "F") > 1
1.024 JMPL J_TYPE_K ZCMPI(S[11], "J") + ZCMPI(S[10], "K") > 1
1.025 JMPL K_TYPE_K ZCMPI(S[11], "K") + ZCMPI(S[10], "K") > 1
1.026 JMPL T_TYPE_K ZCMPI(S[11], "T") + ZCMPI(S[10], "K") > 1
1.027 JMPL E_TYPE_K ZCMPI(S[11], "E") + ZCMPI(S[10], "K") > 1
1.028 JMPL N_TYPE_K ZCMPI(S[11], "N") + ZCMPI(S[10], "K") > 1
1.029 JMPL B_TYPE_K ZCMPI(S[11], "B") + ZCMPI(S[10], "K") > 1
1.030 JMPL R_TYPE_K ZCMPI(S[11], "R") + ZCMPI(S[10], "K") > 1
1.031 JMPL S_TYPE_K ZCMPI(S[11], "S") + ZCMPI(S[10], "K") > 1
1.032 DISP You entered [S11] Type, [S10] Scale for entry.
1.032 DISP This is not a valid entry/combination!
1.032 DISP
1.032 DISP Please enter one of the following types
1.032 DISP Thermcouples: J, K, T, E, N, B, R or S
1.032 DISP Temperature Scale: C, F, K
1.033 JMPL DATA_ENTRY
# Degrees C Scale
1.034 LABEL J_TYPE_C
1.035 DISP J Type in Deg C
# insert J type test here
1.036 JMPL END
1.037 LABEL K_TYPE_C
1.038 DISP K Type in Deg C
# insert K type test here
1.039 JMPL END
1.040 LABEL T_TYPE_C
1.041 DISP T Type in Deg C
# insert T type test here
1.042 JMPL END
1.043 LABEL E_TYPE_C
1.044 DISP E Type in Deg C
# insert T type test here
1.045 JMPL END
1.046 LABEL N_TYPE_C
1.047 DISP N Type in Deg C
# insert T type test here
1.048 JMPL END
1.049 LABEL B_TYPE_C
1.050 DISP B Type in Deg C
# insert T type test here
1.051 JMPL END
1.052 LABEL R_TYPE_C
1.053 DISP R Type in Deg C
# insert R type test here
1.054 JMPL END
1.055 LABEL S_TYPE_C
1.056 DISP S Type in Deg C
# insert B type test here
1.057 JMPL END
# Degrees F Scale
1.058 LABEL J_TYPE_F
1.059 DISP J Type in Deg F
# insert J type test here
1.060 JMPL END
1.061 LABEL K_TYPE_F
1.062 DISP K Type in Deg F
# insert K type test here
1.063 JMPL END
1.064 LABEL T_TYPE_F
1.065 DISP T Type in Deg F
# insert T type test here
1.066 JMPL END
1.067 LABEL E_TYPE_F
1.068 DISP E Type in Deg F
# insert T type test here
1.069 JMPL END
1.070 LABEL N_TYPE_F
1.071 DISP N Type in Deg F
# insert T type test here
1.072 JMPL END
1.073 LABEL B_TYPE_F
1.074 DISP B Type in Deg F
# insert T type test here
1.075 JMPL END
1.076 LABEL R_TYPE_F
1.077 DISP R Type in Deg F
# insert R type test here
1.078 JMPL END
1.079 LABEL S_TYPE_F
1.080 DISP S Type in Deg F
# insert B type test here
1.081 JMPL END
# Degrees K Scale
1.082 LABEL J_TYPE_K
1.083 DISP J Type in Deg K
# insert J type test here
1.084 JMPL END
1.085 LABEL K_TYPE_K
1.086 DISP K Type in Deg K
# insert K type test here
1.087 JMPL END
1.088 LABEL T_TYPE_K
1.089 DISP T Type in Deg K
# insert T type test here
1.090 JMPL END
1.091 LABEL E_TYPE_K
1.092 DISP E Type in Deg K
# insert T type test here
1.093 JMPL END
1.094 LABEL N_TYPE_K
1.095 DISP N Type in Deg K
# insert T type test here
1.096 JMPL END
1.097 LABEL B_TYPE_K
1.098 DISP B Type in Deg K
# insert T type test here
1.099 JMPL END
1.100 LABEL R_TYPE_K
1.101 DISP R Type in Deg K
# insert R type test here
1.102 JMPL END
1.103 LABEL S_TYPE_K
1.104 DISP S Type in Deg K
# insert B type test here
1.105 JMPL END
1.106 LABEL END