Performing high-pressure calibration

Document(s): 
application/pdf icon Performing high-pressure calibration (924.63 KB)

High-pressure measurement is important to ensure the safety of operators. When performing these calibrations, measurement quality and operator safety are critical.

A technician demonstrating the Fluke 8270A

What is high pressure?

High pressure is a relative term. What some consider to be high, others regard as a normal range. Within the scope of this application note, we define high pressure as any pressure greater than 20 MPa (3000 psi). This threshold is higher than what is available in a standard gas cylinder and is an obvious breakpoint when looking at the variety and availability of different types of pressure reference standards. Static pressure calibrations are performed at pressures as high as 500 MPa (75 000 psi), with rare applications requiring pressures as high as 1 GPa (150 000 psi). This application note focuses on the lower end of this high-pressure regime, but many of the considerations discussed can be extrapolated up to these very high-pressures.

The physics behind pressure calibration

To examine all of the intricacies of high-pressure calibration, we must first understand the physics involved and why the system works the way it does.

Ideal Gas Law approximates the behavior of many gasses under different conditions, but with several limitations. For our purposes, it is useful because it describes the variables that affect the pressure in an enclosed system. Specifically, it states:

PV=nRT

Where:

P = Pressure

V = Volume

n = Number of molecules

R = Ideal gas constant

T = Temperature

This tells us that pressure is impacted by three things:

  • Volume of the system
  • Number of molecules in the system
  • Temperature of the fluid

The relationship between pressure and volume is inverse; as the volume decreases, the pressure increases. The number of molecules and the temperature have a direct relationship. What does this mean?

To change the pressure in a system, there are three things that can be done:

  • Change the volume. An example of this is a screw press or variable volume on a pressure comparator or deadweight tester.
  • Change the number of molecules in a system. This is how most gas pressure controllers operate. Additional gas is metered in or exhausted out of the system to increase or decrease the pressure.
  • Change the temperature in order to change the pressure. This is a rarely used approach but can be used to stabilize the pressure at a given set point.

The Ideal Gas Law is also useful in understanding why the pressure is unstable. Changes in pressure must align with one or more of the three variables: volume change (expansion or contraction of pressure lines); change in number of molecules (a leak); or temperature effects (as the temperature of the fluid cools, the pressure decreases with it).

Viscosity

Viscosity is the property of a fluid to resist movement or flow. Theoretically, when measuring static pressure, the media should be static and viscosity should not have a significant effect. When working with high pressure, it’s not always that simple. For safety reasons, small tubing is often used for high pressure. If a viscous fluid (like some oils) is used, this could result in excessive time for pressure stability. This is especially important to consider when testing will be performed at different temperatures, as the viscosity of a fluid changes as the temperature changes.

The Fluke 8270A High Pressure Controller
The Fluke 8270A High Pressure Controller

Compressibility

Compressibility is the property of a fluid that dictates how much its volume will change when a pressure is applied. The Ideal Gas Law has already told us the pressure and volume of the system are highly linked. How much the pressure will change with a volume change is defined by the fluid’s compressibility. A gas is highly compressible. If the volume of the system changes, the pressure will not dramatically change with it. Conversely, liquids are considered incompressible. Therefore, reducing the volume of the system will dramatically increase the pressure. This makes it easier to generate high pressures with liquids than with gas.

Head pressure

The weight of a column of fluid will generate a pressure on the bottom of the column. An example of this is a swimming pool; the water pressure is greater at the bottom of the pool than it is on the surface. Mathematically, this is shown in this equation:

P=hgρ

P = Pressure (absolute)

h = height of the fluid column

g = acceleration of gravity

ρ = density of the fluid

If we are measuring the pressure in a system at two different places (reference and test pressures) and these points are at two different vertical heights, then there will be a difference in the pressure caused by the head pressure. If the pressure medium is a liquid, the density of the fluid is considered a constant with pressure and thus the head pressure is a constant, regardless of the overall pressure of the system. If the pressure medium is a gas, the density is lower but changes with pressure.

At low pressures, a liquid head pressure can be significant in the overall pressure measurement. At higher pressures, the head pressure becomes less significant compared to the overall pressure.

Considerations when performing high-pressure calibrations

Safety

The first thing to consider when performing high-pressure calibrations is safety. This includes:

  • Proper personal protection equipment (PPE), such as the use of safety glasses
  • Proper selection of equipment. All pressure boundary equipment (including items like fittings, tubing, valves, and gauges) must be rated for the appropriate maximum working pressure.
  • Operator training. The user of the calibration equipment needs to be properly trained in the safe operation of the equipment. This is especially true for manually operated equipment that requires the operator to physically turn screw presses or valves.
  • Inclusion of pressure limiting devices like relief valves or burst discs provide a last defense against an over-pressure situation and sensors with large transmitter heads need calibration. When analyzing the tank size required for your application, be sure to allow for proper sensor immersion depth and fluid space below the sensors and between the sensors being tested and the tank wall.

Fluid selection

What medium should be used when performing pressure calibrations? At the extreme ends of the pressure range, the choices are more obvious. For lower pressures, especially pressures below 7 MPa (1000 psi), a gas medium is most common. The pressures are relatively easy to generate, gas is clean, and it eliminates large head pressure errors. For pressures above 100 MPa (15 000 psi), there are very few equipment choices for using a gas medium and generating the pressure gets difficult. Oil should be used, unless there is a specific compatibility issue that makes it impossible.

What about all the pressures in between? In calibration laboratory applications, gas media have been commonly used at pressures up to (or even above) 14 MPa (2000 psi) where nitrogen cylinders are readily available. In recent years, the breakpoint between gas and liquid media selection has continued to rise as higher pressure gas cylinders or means for boosting to higher gas pressures become available. The advantages to using gas instead of liquid include a reduced chance of contaminating the device under test (DUT); improved stability time due to lower viscosity; reduction of head pressure errors; and improved options for automated equipment. These advantages have to be weighed against the disadvantages, which include the fact that high-pressure gas has more compressed energy, making it more dangerous if an over-pressure event occurs. Following the safety guidelines given above will help mitigate this risk.

Fluid compatibility

An important consideration in selecting the fluid is to ensure that it is compatible with the devices that will be tested as well as the reference standard. This is no easy task. Most high-pressure measurement devices are used with a liquid medium in their “real world application.” If your calibration system uses a gas medium, then the liquid from the DUT could contaminate the system. One way to eliminate this issue is to use a contamination prevention system. It allows a liquid-filled device to be calibrated using a gas pressure controller without contaminating the pressure controller. Another approach is to use a reference standard that uses liquid as the medium. In this situation, there is still a concern if the reference standard and the DUT use different liquids. For example, a deadweight tester or piston gauge is designed to only operate with a specific fluid. If a different oil than the one provided is used, it may damage or alter the behavior of the piston-cylinder assembly. One workaround is to use a liquid-to-liquid separator.

Pressure stability

To measure pressure properly, the pressure must be sufficiently stable. This can be challenging at high pressures. As discussed previously, the pressure will be unstable due to changes in volume (expansion and contraction of pressure tubing), leaks, and temperature effects. Of these, the biggest challenge is temperature effects. As the pressure in a system increases, the temperature of the fluid also increases. When the pressure is no longer increasing, the temperature will decrease until it reaches equilibrium. As the temperature decreases, the pressure decreases with it. This decrease in pressure may look like a leak, even though no leak exists. A pressure decrease caused by a leak will continue to decrease in pressure with no settling or slowing down. A pressure decrease caused by temperature will eventually settle out if given sufficient time to stabilize. How do we overcome these temperature effects?

  • Using a deadweight tester or piston gauge. The floating piston acts as a natural regulator to maintain a stable pressure. As the temperature decreases, the piston sinks, changing the volume of the system. As long as the piston is floating with no friction between the piston and cylinder, the system will be at equilibrium and the pressure will be stable.
  • Using a pressure controller. An automated pressure controller actively attempts to maintain a stable pressure by counteracting the temperature effects, either adjusting the number of molecules in the system or changing the volume in the system.
  • Using a pressure comparator or similar manual device. This is where temperature effects are most noticeable, as there is nothing to counteract them. For applications where hysteresis is not an issue, increase the pressure past the setpoint and then decrease the pressure. This will balance out the temperature effects and cause the device to stabilize quicker. If this is not feasible, then extended stabilization time may be necessary.

Pressure Calibrator Fittings and Tubing

Fitting and tubing selection

When setting up a high-pressure calibration system, there are important considerations when selecting the fittings and tubing to be used. The first is that everything must be properly rated for the pressures in question. All pressure fittings and tubing will be rated for a given maximum working pressure by the manufacturer. It should only be used within the rated pressures. With that in mind, it may seem prudent to get the highest possibly rated pressure tubing and always use it, no matter what pressure you are actually going to generate. There are disadvantages to this approach. Higher pressure tubing normally has a smaller internal diameter. When used with viscous fluids this will increase settling times.

A balance has to be established that ensures proper pressure ratings and flow rates. Care should also be taken when using flexible tubing at high pressures. A pressurized unsecured tube can whip around, causing damage or injury. Make sure all fittings are tight and the system is secure before pressurizing.

High-pressure reference standards

There are a number of different types of reference standards available for high-pressure calibration, each with their own advantages and disadvantages.

Equipment type Advantages Disadvantages Example
Hydraulic comparators Low cost Cannot be automated
Slow to stabilize pressure
Fluke P5515 hydraulic comparator
Hydraulic controllers Can be automated
Actively maintains the pressure
Expensive
Large and bulky
Pressure is less stable than pneumatic
controllers
Does not control low pressures well
Fluke PPCH hydraulic controller
Hydraulic deadweight testers Low cost
Wide pressure range coverage
Piston-cylinder provides stable
pressures
Good long term stability provides
longer calibration cycles
Must be corrected for local gravity
Cannot be automated
Limited capabilities at low pressure
Fluke P3124 hydraulic deadweight tester
Hydraulic piston gauges Extremely accurate
Can be automated
Can handle very high pressures
Expensive
Potential contamination of DUT when calibrating
gas-filled devices
Fluke PG7302 hydraulic piston gauges
Pneumatic controllers Medium cost
Wide pressure range coverage
Can be automated
Simple to use
Can be used to calibrate devices with
variety of media (with CPS)
Requires external gas supply
More expensive than hydraulic comparators or
hydraulic deadweight testers
Fluke 8270A pneumatic controller
Pneumatic piston gauges Extremely accurate
Can be automated
Expensive
Longer stabilization times than controllers
Fluke PG7302 pneumatic piston gauge

  • 홈
  • 제품
    • 신제품
    • 전기 교정
      • 전기 표준기
      • 전기 교정기
      • 벤치형 멀티미터
      • 전기 교정 소프트웨어
    • RF 교정
      • RF 레퍼런스 소스
      • RF 교정 액세서리
      • RF 교정 소프트웨어
    • 데이터 수집 및 테스트 장비
      • 벤치 멀티미터
      • 데이터 수집
      • 데이터 수집 소프트웨어
    • 온도 교정
      • ITS-90 온도 표준
        • ITS-90 고정점 셀
        • 표준 백금 저항 온도계(SPRT)
        • 유지장치
        • 액화 질소 비교 교정기
        • 저항 브리지
        • 표준 저항기
      • 교정 항온조
        • 소형 교정 항온조
        • 표준 교정 항온조
        • 특수 응용 항온조
        • 교정 항온조 액세서리
        • 교정 항온조 교정기
        • 교정 항온조 매질
      • 산업 온도 교정기
        • 현장용 계측 웰
        • 계측 웰
        • 휴대용 드라이 웰 교정기
        • 현장 드라이 블록 교정기
        • 마이크로 항온조
        • 적외선 교정기
        • 열전대 전기로
        • 이중 블록 드라이 웰
        • 영점 드라이-웰
      • 프로브/센서
        • 백금 저항 온도계(PRT)
        • 서미스터
        • 열전대
      • 디지털 온도계 리드아웃
      • 다기능 교정기
      • 온도 교정 소프트웨어
    • 습도 교정
      • 습도 발생기
      • 습도 데이터 로거 및 모니터
    • 압력 교정
      • 피스톤게이지
        • NMI 피스톤 게이지
        • 절대압 피스톤 게이지
        • 고압 압축 공기 피스톤 게이지
        • 유압 피스톤 게이지
        • 피스톤 게이지 엑세서리
      • 압력 컨트롤러/교정기
        • 저압력 제어기/ 교정기
        • 공기압식 압력 제어기/교정기
        • 고압의 공압 제어기/교정기
        • 유압 제어기/교정기
      • 압력 모니터
        • 저압력 모니터
        • 디지털 압력 게이지
        • 레퍼런스 압력 모니터
      • 분동식 압력계
        • 공압 분동식 테스트기
        • 유압 분동식 테스터기
        • 수압 분동식 테스터기
        • 고압력 유압 분동식 테스트기
        • 분동식 압력 테스터기 액세서리
      • 수동 압력 교정기 및 모니터
        • 압력 교정기
        • 공압식 압력 제어기
        • 유압 압력 비교 측정기/펌프
      • Handheld Pressure Calibrators
      • 대기 데이터 교정
      • 환경 모니터링
      • 압력 교정 액세서리
      • 압력 교정 커스텀 시스템
      • 압력 교정 소프트웨어
    • 유량 교정
      • 기체 유량 표준기
      • 기체 유량 액세서리
      • GFS 일차 표준 유량 표준기
      • 유량 교정 소프트웨어
    • 프로세스 교정 장비
      • 온도 교정기
        • 휴대용 온도 교정기
        • 드라이 블록 교정기 및 초소형 항온조
        • 정밀 디지털 온도계
        • 온도 프로브
        • 적외선 교정기
        • 데이터 로깅 기능이 있는 온습도계
      • 압력 교정기
        • 디지털 압력 교정기
        • 휴대용 압력 교정기
        • 분동식 압력계
        • 정밀 디지털 압력 게이지
        • 교정 핸드 펌프
      • 다기능 교정기
      • mA 루프 교정기
      • 프로세스 교정 소프트웨어
    • 교정 소프트웨어
      • MET/CAL® 소프트웨어
      • MET/CAL® 지원
      • 자산 관리 소프트웨어
      • 온도 교정 소프트웨어
      • 압력 교정 소프트웨어
      • 유량 교정 소프트웨어
      • 범용 교정 소프트웨어
    • 서비스 및 지원
    • 모든 교정 기기
  • 구입처 안내
    • 구입처
    • 견적 요청
    • 데모 요청
    • 구입 상담 요청
    • Certified Pre-Owned Equipment
    • General Services Administration (GSA)
    • Financing Program
    • National Stock Numbers (NSNs)
    • Payment Options and Tax Info
  • 뉴스
    • 보도 자료
    • 프로모션
    • Metrology Salary Survey
  • 교육 및 세미나
    • 컨퍼런스 및 전시회
    • 교육 과정
    • User Group Meetings
    • 웹세미나
      • 실시간 세미나
      • 보관된 세미나
  • 사용자 가이드
    • Education Hub
    • About Calibration
    • 문서 및 교육
    • 제품 자료
    • 제품 매뉴얼(사용자 가이드)
    • 동영상 및 가상 데모
    • Blog
  • 서비스 및 지원
    • Service Request (RMA)
    • Service Plans
    • 기술 자료
    • 제품 인증
    • Calibration Certificates
    • 교정 커뮤니티 사이트
    • My MET/SUPPORT
      • Activate
      • Procedures
      • Software
      • Technical Bulletins
      • Priority Support
      • Manuals
    • 제품 매뉴얼(사용자 가이드)
    • 안전 보건 자료(SDS)
    • Safety, Service, and Product Notices
    • 소프트웨어 다운로드
  • 회사 정보
    • 연락처
    • 회사 위치
    • Fluke Calibration
      • Hart Scientific 온도 교정
      • DHI 압력 및 유량 교정
      • Pressurements 압력 교정
      • Ruska 압력 교정
    • 채용 정보
    • 자회사 안내
    • 도덕성 및 규정 준수
    • FAQ
홈|자료|문서 및 교육|압력 교정|어플리케이션 노트|Performing high-pressure calibration

©1995-2023 Fluke Corporation
             

Secondary menu

  • Disclaimer
  • 개인정보 취급방침
  • 이용약관
  • 판매 약관